![]() Date 03-APR-14Ģ Optional Steps Once the first i-stat 1 Handheld has been updated using the JammLite Utility, additional i-stat 1 Handhelds may be updated the same way or by using the Handheld-to-Handheld method. This process is noted in the diagram below. ![]() JammLite Process Overview Whether updating one, two, or many i-stat 1 Handheld(s), the JammLite procedure must be used to update the first i-stat 1 Handheld. New JAMS application software allows the i-stat 1 Handheld to recognize any newly launched cartridge types and to perform any newly launched features. New CLEW software delivered twice a year on a CD-ROM re-establishes these standardization values and incorporates refinements to the internal quality monitoring system. These updates are equivalent to manually adjusting calibration on a traditional laboratory analyzer. Due to the continuous manufacturing process improvements to the i-stat System, it is necessary to update standardization values from time to time to maintain long-term consistency of performance. Why Do i-stat Handhelds Need Software Updates? The i-stat 1 System is designed to eliminate operator influence on delivered results. This process should take less than 10 minutes for the first Handheld, and less time for additional Handhelds. 1 i-stat TECHNICAL BULLETIN INSTRUCTIONS FOR UPDATING i-stat 1 HANDHELD SOFTWARE Overview This Technical Bulletin has been created specifically to guide you through the process of updating the software on your i-stat 1 Handheld(s).